Verify Calibrator Settings for GE Druck

Calibrator settings are saved for a user and for a data source. If you log in as another user, or change the data source, you must reconfigure the calibrator settings.

Steps

  1. Access the Calibration Management Overview page.

    Hint: You can also access the Calibration Management Overview page for a specific hierarchy level.

  2. Select the Calibration Tools tab.

    The Calibration Tools section appears.

  3. In the upper-right corner of the section, select Calibrator Settings.

    The Calibrator Settings window appears.

    Calibration Settings

  4. In the Select Device box, select one of the following calibrators:

    The calibration settings for the selected calibrator appear.

    The Calibration Settings window

  5. In the Select Serial Number box, select the serial number of the GE Druck calibrator.

    The Done button is enabled.

  6. If you want to test the connection of the calibrator, select the Perform Connection Test check box.
  7. In the Device Service Settings Service Port box, enter the value of the service port number that you have configured. The default value in the Device Service Settings Service Port box is 2014.

    The Done button is enabled.

  8. In the lower-right corner of the window, select Done.

    Note: If you selected the Perform Connection Test check box, then the connection between GE Digital APM and the calibrator is tested.
    • If you did not install the Meridium Device Service, then a message appears, asking you to install it. You must install the Meridium Device Service, and then select Continue.
    • If the connection is successful, then the details appear in the Device Log box.

    The calibration settings are saved.

What's Next?

Copyright © 2018 General Electric Company. All rights reserved.