About Calibration Event Graphs

Calibration event graphs provide a visual representation of the data that exists in the Calibration Event records. By default, these graphs are stored in the following Catalog folder: \\Public\Meridium\Modules\Calibration Management\Graphs.

The following table lists the types of graphs in Calibration Management:

Name of the graph Description Types of calibration that the graph is applicable to
Error Graph Plots the As Found and As Left error values for each of the test points. The graph also shows the upper and lower limits for the error as specified in the associated calibration template.
  • Analog
  • Weight Scale
  • Single Component Analyzer
  • Multi-Component Analyzer
Measure Graph Plots the As Found and As Left output values for each of the test points.
  • Analog
  • Weight Scale
  • Single Component Analyzer
Error Trend Plots the As Found and As Left errors for each calibration event for a selected period of time.
  • Analog
  • Discrete
  • Weight Scale
  • Single Component Analyzer
  • Multi-Component Analyzer
Pass/fail Trend Plots the number of passed, failed, and total calibration events for a selected period of time.
  • Analog
  • Discrete
  • Weight Scale
  • Single Component Analyzer
  • Multi-Component Analyzer

Copyright © 2018 General Electric Company. All rights reserved.